Thiamine-reineckate liquid membrane electrode for the selective determination of thiamine (vitamin B1) in pharmaceutical preparations
المؤلف | Hassan, Saad S.M. |
المؤلف | Elnemma, Eman M. |
تاريخ الإتاحة | 2024-03-04T08:58:29Z |
تاريخ النشر | 1989 |
اسم المنشور | The Analyst |
المصدر | Scopus |
الرقم المعياري الدولي للكتاب | 0003-2654 |
الملخص | The construction and electrochemical response characteristics of a liquid membrane electrode for thiamine (vitamin B1) based on the formation of an ion pair between the thiamine cation and the reineckate anion in nitrobenzene are described. The electrode showed a linear response for thiamine over the concentration range 10–3–10–6–M with a cationic slope of 31.6–mV per concentration decade and had a wide working pH range (4–8), fast response time (1–2–min), low detection limit (0.3–µg–ml–1) good stability and reasonable selectivity. The direct potentiometric determination of vitamin B1 in some pharmaceutical preparations using the proposed electrode gave an average recovery of 98.1% of the nominal values and a mean standard deviation of 1.3%, which compare fairly well with data obtained using the British and United States pharmacopoeial methods. |
اللغة | en |
الناشر | The Royal Society of Chemistry |
الموضوع | Liquid membrane electrode Pharmaceutical analysis Reineckate Thiamine-selective electrode Vitamin B-, determination |
النوع | Article |
الصفحات | 735-737 |
رقم المجلد | 114 |
الملفات في هذه التسجيلة
الملفات | الحجم | الصيغة | العرض |
---|---|---|---|
لا توجد ملفات لها صلة بهذه التسجيلة. |
هذه التسجيلة تظهر في المجموعات التالية
-
الكيمياء وعلوم الأرض [587 items ]