AbstractDuring cystoscopy, the urologist manually steers a cyst scope inside a patient's bladder to visually inspect the inner surface. Cystoscopies are performed as part of surveillance for bladder cancer, making it the most expensive cancer to treat over a patient's lifetime. An automated bladder scanning system has been devised to reduce workload and cost by relieving the urologist from performing surveillance. Presented here is a proof-of-concept apparatus that controls the motion of a miniature flexible endoscope. Image-based feedback is used to adjust the endoscope's movement so that captured images overlap with one another, ensuring that the entire inner surface of the bladder is imaged. Within a bladder phantom, the apparatus adaptively created and followed a spherical scan pattern comprised of 13 individual latitudes and 508 captured images, while accepting between 60% and 90% image overlap between adjacent images. The elapsed time and number of captured images were sensitive to the apparatus's placement within the phantom and the acceptable image overlap percentage range. A mosaic of captured images was generated to validate comprehensive surveillance. Overall, a robotically controlled endoscope used in conjunction with image-based feedback may permit fully automated and comprehensive bladder surveillance to be conducted without direct clinician oversight.
SponsorNational Priorities Research Program (NPRP) under Grant NPRP 09-214-2-090 from the Qatar National Research Fund (a member of Qatar Foundation).
Languageen
PublisherIEEE
SubjectAdaptive control
bladder
endoscope
robotic
TitleControlling the trajectory of a flexible ultrathin endoscope for fully automated bladder surveillance
TypeArticle
Pagination366-373
Issue Number1
Volume Number19
dc.accessType Abstract Only


Files in this item

FilesSizeFormatView

There are no files associated with this item.

This item appears in the following Collection(s)

Show simple item record