Characterization of capacitive electromyography biomedical sensor insulated with porous medical bandages
عرض / فتح
التاريخ
2020المؤلف
Ng, Charn LoongReaz, Mamun Bin Ibne
Crespo, Maria Liz
Cicuttin,
res
Chowdhury, Muhammad Enamul Hoque
...show more authors ...show less authors
البيانات الوصفية
عرض كامل للتسجيلةالملخص
A capacitive electromyography (cEMG) biomedical sensor measures the EMG signal from human body through capacitive coupling methodology. It has the flexibility to be insulated by different types of materials. Each type of insulator will yield a unique skin-electrode capacitance which determine the performance of a cEMG biomedical sensor. Most of the insulator being explored are solid and non-breathable which cause perspiration in a long-term EMG measurement process. This research aims to explore the porous medical bandages such as micropore, gauze, and crepe bandage to be used as an insulator of a cEMG biomedical sensor. These materials are breathable and hypoallergenic. Their unique properties and characteristics have been reviewed respectively. A 50 Hz digital notch filter was developed and implemented in the EMG measurement system design to further enhance the performance of these porous medical bandage insulated cEMG biomedical sensors. A series of experimental verifications such as noise floor characterization, EMG signals measurement, and performance correlation were done on all these sensors. The micropore insulated cEMG biomedical sensor yielded the lowest noise floor amplitude of 2.44 mV and achieved the highest correlation coefficient result in comparison with the EMG signals captured by the conventional wet contact electrode. 2020, The Author(s).
المجموعات
- الهندسة الكهربائية [2649 items ]