Integration technique for neutron source standardization inside A cylindrical water tank
المؤلف | El Kolaly, M. A. [محمد عبد الفتاح القللي] |
تاريخ الإتاحة | 2009-11-25T15:16:09Z |
تاريخ الإتاحة | 2015-01-19T05:27:23Z |
تاريخ النشر | 1993 |
اسم المنشور | Qatar University Science Journal |
الاقتباس | Qatar University Science Journal, 1993, Vol. 13, No. 1, Pages 14-17. |
الملخص | This work deals with the evaluation of neutron irradiation facility. Two (Pu-Be) sources and two (Ra-Be) sources are used. Integration technique by a BF3 detector for neutron source standardization is used. The necessary calibrations have been performed, and all the possible known errors have been taken into consideration. The total error using this technique does not exceed ± 10%. This assembly is capable of safely accommodating the neutron sources and of permitting work inside the assembly without any hazard. Moreover, the availability of such neutron irradiation facilities with isotropic neutron sources provide a promising tool for applications to neutron activation analysis. |
اللغة | en |
الناشر | Qatar University |
الموضوع | Physics الفيزياء |
العنوان البديل | معايرة المصادر النيوترونية بطريقة المسح الاشعاعي الكلي داخل وعاء اسطواني مملوء بالماء |
النوع | Article |
الصفحات | 14-17 |
رقم العدد | 1 |
رقم المجلد | 13 |
الملخص البديل | يتناول هذا البحث تقدير الانبعاث النيوتروني من المصادر النيوترونية - ولقد استخدمت في هذه القياسات أربعة مصادر مختلفة الطاقة - كما استخدم في المسح النيوتروني الكلي الكاشف (ثلاثي فلوريد البورون – BH3) . ولقد قدرت كمية النشاط الاشعاعي لهذه المصادر كما حسبت نسبة الخطأ الكلي التي وجد أنها لا تزيد عن ±10% ، وتتميز هذه الطريقة بأنها لا تشكل أية خطورة من ناحية التعرض الاشعاعي ، بالإضافة إلى امكانية استخدامها في أعمال التنشيط الاشعاعي ، وكذلك يمكن تطبيقها في أعمال اشعاعية أخرى . |
الملفات في هذه التسجيلة
هذه التسجيلة تظهر في المجموعات التالية
-
مجلة جامعة قطر للعلوم - [من 1981 الى 2007] [770 items ]