Show simple item record

AuthorKristensen, Søren Risom
AuthorNybo, Jette
AuthorPedersen, Shona
Available date2022-04-24T06:56:14Z
Publication Date2022-01-01
Publication NameResearch and Practice in Thrombosis and Haemostasis
Identifierhttp://dx.doi.org/10.1002/rth2.12654
CitationKristensen, SR, Nybo, J, Pedersen, S. Thrombin generation measured on ST Genesia, a new platform in the coagulation routine lab: Assessment of analytical and between-subject variation. Res Pract Thromb Haemost. 2022; 6:e12654. doi:10.1002/rth2.12654
URIhttps://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=85124456047&origin=inward
URIhttp://hdl.handle.net/10576/30241
AbstractBackground: The thrombin generation (TG) assay, which measures global coagulation, has mainly been used as a research tool to investigate thrombotic and bleeding disorders. Recently, Diagnostica Stago launched the ST Genesia, a fully automated system to perform “routine version” of this assay. The objectives of this study were to evaluate the imprecision compared with the previous method, Thrombinoscope CAT, and to establish reference intervals. Methods: Thrombin generation was measured in platelet-poor citrated plasma from 20 normal controls (fresh and after freezing at −80°C up to 12–13 weeks) on CAT and ST Genesia in duplicate to estimate the total variation, and within and between variations. The reference intervals were estimated nonparametrically in 30 men, 30 women taking combined oral contraceptives (COCs), and 30 women not taking COCs. These were sampled in both Vacutainer and Monovette tubes (i.e., tubes with a high and minimal contact activation, respectively). Results: Freezing had minimal effects. Imprecision was comparable between the ST Genesia and CAT, with a strong correlation between the two methods. TG was higher when sampled in Vacutainer than in Monovette. We observed a distinct difference between women taking and not taking COCs, whereas men and women not taking COC were quite similar. Conclusions: Thrombin generation on ST Genesia showed an analytical variation similar to that of CAT. The results depended on the type of sample tubes; thus, reference intervals must be established for the collection tubes used in each laboratory. Furthermore, a considerable difference was observed between women using and not using COCs.
Languageen
PublisherWiley Open Access
SubjectCAT
contact activation
imprecision
reference intervals
ST genesia
thrombin generation
TitleThrombin generation measured on ST Genesia, a new platform in the coagulation routine lab: Assessment of analytical and between-subject variation
TypeArticle
Issue Number1
Volume Number6
ESSN2475-0379
dc.accessType Open Access


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record