Motion Artifacts Correction from Single-Channel EEG and fNIRS Signals Using Novel Wavelet Packet Decomposition in Combination with Canonical Correlation Analysis
عرض / فتح
التاريخ
2022المؤلف
Hossain, Md S.Chowdhury, Muhammad E. H.
Reaz, Mamun B.
Ali, Sawal H.
Bakar, Ahmad Ashrif A.
Kiranyaz, Serkan
Khandakar, Amith
Alhatou, Mohammed
Habib, Rumana
Hossain, Muhammad M.
...show more authors ...show less authors
البيانات الوصفية
عرض كامل للتسجيلةالملخص
The electroencephalogram (EEG) and functional near-infrared spectroscopy (fNIRS) signals, highly non-stationary in nature, greatly suffers from motion artifacts while recorded using wearable sensors. Since successful detection of various neurological and neuromuscular disorders is greatly dependent upon clean EEG and fNIRS signals, it is a matter of utmost importance to remove/reduce motion artifacts from EEG and fNIRS signals using reliable and robust methods. In this regard, this paper proposes two robust methods: (i) Wavelet packet decomposition (WPD) and (ii) WPD in combination with canonical correlation analysis (WPD-CCA), for motion artifact correction from single-channel EEG and fNIRS signals. The efficacy of these proposed techniques is tested using a benchmark dataset and the performance of the proposed methods is measured using two well-established performance matrices: (i) difference in the signal to noise ratio (ΔSNR
) and (ii) percentage reduction in motion artifacts (η
). The proposed WPD-based single-stage motion artifacts correction technique produces the highest average ΔSNR
(29.44 dB) when db2 wavelet packet is incorporated whereas the greatest average η
(53.48%) is obtained using db1 wavelet packet for all the available 23 EEG recordings. Our proposed two-stage motion artifacts correction technique, i.e., the WPD-CCA method utilizing db1 wavelet packet has shown the best denoising performance producing an average ΔSNR
and η
values of 30.76 dB and 59.51%, respectively, for all the EEG recordings. On the other hand, for the available 16 fNIRS recordings, the two-stage motion artifacts removal technique, i.e., WPD-CCA has produced the best average ΔSNR
(16.55 dB, utilizing db1 wavelet packet) and largest average η
(41.40%, using fk8 wavelet packet). The highest average ΔSNR
and η
using single-stage artifacts removal techniques (WPD) are found as 16.11 dB and 26.40%, respectively, for all the fNIRS signals using fk4 wavelet packet. In both EEG and fNIRS modalities, the percentage reduction in motion artifacts increases by 11.28% and 56.82%, respectively when two-stage WPD-CCA techniques are employed in comparison with the single-stage WPD method. In addition, the average ΔSNR
also increases when WPD-CCA techniques are used instead of single-stage WPD for both EEG and fNIRS signals. The increment in both ΔSNR
and η
values is a clear indication that two-stage WPD-CCA performs relatively better compared to single-stage WPD. The results reported using the proposed methods outperform most of the existing state-of-the-art techniques.
المجموعات
- الهندسة الكهربائية [2649 items ]